Pipette Performance Check (Demo)
published Version 1 11/20/2026, 2:42:54 PM
Initial release
Document
Pipette Performance Check (Demo)
1. Purpose
This procedure describes a routine gravimetric check of air-displacement pipettes between calibrations.
2. Scope
Single-channel adjustable pipettes (P10 to P1000) used in the lab.
Disclaimer: This is a demonstrational document only and should not be used for actual laboratory work without verification by qualified personnel.
3. Responsibilities
- Each user checks their assigned pipettes once per month.
- The lab manager reviews the log and arranges service for failed pipettes.
4. Materials and Equipment
- Analytical balance (readability 0.1 mg or better)
- Weighing vessel with lid
- Deionized water at room temperature
- Thermometer
- Matching pipette tips
5. Procedure
- Equilibrate: Keep water, pipette, and tips in the room for at least 1 hour. Record the water temperature.
- Tare: Place the weighing vessel on the balance and tare it.
- Pre-wet: Aspirate and dispense the test volume three times to wet the tip.
- Weigh: Dispense the nominal volume into the vessel and record the mass. Repeat for 10 readings.
- Calculate: Convert mass to volume with the density of water at the recorded temperature.
- Assess: Compare the mean volume (accuracy) and standard deviation (precision) to the limits in Table 1.
| Nominal volume | Accuracy limit | Precision limit |
|---|---|---|
| 10 µL | ± 1.0 % | ≤ 0.8 % |
| 100 µL | ± 0.8 % | ≤ 0.3 % |
| 1000 µL | ± 0.8 % | ≤ 0.3 % |
Table 1: Example acceptance limits for this demo.
6. Results
- Pass: Record the result in the pipette log and attach a dated check label.
- Fail: Label the pipette "Do not use", remove it from service, and inform the lab manager.
7. Records
Keep the pipette log for at least three years.