Pipette Performance Check (Demo)

published Version 1 11/20/2026, 2:42:54 PM

Initial release

Document

Pipette Performance Check (Demo)

1. Purpose

This procedure describes a routine gravimetric check of air-displacement pipettes between calibrations.

2. Scope

Single-channel adjustable pipettes (P10 to P1000) used in the lab.

Disclaimer: This is a demonstrational document only and should not be used for actual laboratory work without verification by qualified personnel.

3. Responsibilities

  • Each user checks their assigned pipettes once per month.
  • The lab manager reviews the log and arranges service for failed pipettes.

4. Materials and Equipment

  1. Analytical balance (readability 0.1 mg or better)
  2. Weighing vessel with lid
  3. Deionized water at room temperature
  4. Thermometer
  5. Matching pipette tips

5. Procedure

  1. Equilibrate: Keep water, pipette, and tips in the room for at least 1 hour. Record the water temperature.
  2. Tare: Place the weighing vessel on the balance and tare it.
  3. Pre-wet: Aspirate and dispense the test volume three times to wet the tip.
  4. Weigh: Dispense the nominal volume into the vessel and record the mass. Repeat for 10 readings.
  5. Calculate: Convert mass to volume with the density of water at the recorded temperature.
  6. Assess: Compare the mean volume (accuracy) and standard deviation (precision) to the limits in Table 1.
Nominal volume Accuracy limit Precision limit
10 µL ± 1.0 % ≤ 0.8 %
100 µL ± 0.8 % ≤ 0.3 %
1000 µL ± 0.8 % ≤ 0.3 %

Table 1: Example acceptance limits for this demo.

6. Results

  • Pass: Record the result in the pipette log and attach a dated check label.
  • Fail: Label the pipette "Do not use", remove it from service, and inform the lab manager.

7. Records

Keep the pipette log for at least three years.